Description
RESIDUAL CHLORINE TEST KIT,COLOUR COMPARATOR,
Product Technical bulletin
Chlorine Test Kit Using Procedure
Rxsol-62-5517-020
Range : 1-25 ppm And 26 – 50 ppm
Direction For Use :
1. Take 10 ml of water sample to be tested in a test tube.
2. Add 4 drop of Reagent TK CL 13 and mix well
3. Add 7 drop of Reagent TK CL 14. Mix well
4. Add 3 drop of Reagent TK CL 15. Mix well
5. Now drop wise add Reagent TK CL 16. Counting the number of drop while mixing, until colour changes from BLUE TO COLOURLESS.
# If the expected ppm level of chlorine is more than 20 PPM then use REAGET TK CL17
CALCULATION :
Chlorine ppm as Cl2= 1 X Number of drops of Reagent TK CL 16
Chlorine ppm as Cl2= 10 X Number of drops of Reagent TK CL 17
Product label